Repository logo
 

Automated, small sample-size adiabatic calorimeter.

dc.contributor.authorVan Oort, Michiel J. M.en_US
dc.contributor.authorWhite, Mary Anneen_US
dc.date.accessioned2013-08-23T15:58:43Z
dc.date.available2013-08-23T15:58:43Z
dc.date.issued1987en_US
dc.description.abstractAn automated adiabatic calorimeter is described with an internal vol. of 5 cm3, operable at 30-380 K. One of the main advantages of this calorimeter over others in use is the shorter down time during sample changes, due to interchangeable sample vessels that fit into the heater/thermometer assembly. This calorimeter was tested by measuring the heat capacity of benzoic acid, and the results agreed with the literature values to within 0.5%.en_US
dc.identifier.citationVan Oort, Michiel J. M., and Mary Anne White. 1987. "Automated, small sample-size adiabatic calorimeter.." Review of Scientific Instruments 58(7): 1239-1241.en_US
dc.identifier.issn0034-6748en_US
dc.identifier.issue7en_US
dc.identifier.startpage1239en_US
dc.identifier.urihttp://dx.doi.org/10.1063/1.1139445en_US
dc.identifier.urihttp://hdl.handle.net/10222/36221
dc.identifier.volume58en_US
dc.relation.ispartofReview of Scientific Instrumentsen_US
dc.titleAutomated, small sample-size adiabatic calorimeter.en_US
dc.title.alternativeSection Title: General Physical Chemistryen_US
dc.typearticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Van Oort_Automated.pdf
Size:
462.45 KB
Format:
Adobe Portable Document Format
Description:

Collections