Repository logo
 

Automated, small sample-size adiabatic calorimeter.

Date

1987

Authors

Van Oort, Michiel J. M.
White, Mary Anne

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

An automated adiabatic calorimeter is described with an internal vol. of 5 cm3, operable at 30-380 K. One of the main advantages of this calorimeter over others in use is the shorter down time during sample changes, due to interchangeable sample vessels that fit into the heater/thermometer assembly. This calorimeter was tested by measuring the heat capacity of benzoic acid, and the results agreed with the literature values to within 0.5%.

Description

Keywords

Citation

Van Oort, Michiel J. M., and Mary Anne White. 1987. "Automated, small sample-size adiabatic calorimeter.." Review of Scientific Instruments 58(7): 1239-1241.